Multifunction Process Calibrator
SPMK3005 Multifunction Process Calibrator (also named as: Thermometer Calibrator) is customized for calibration of thermodynamic instruments, pressure gauges etc.
Features:
lSource current, voltage, resistance, frequency, pressure.
lSimulate electrical signals output of thermocouple and thermal resistors.
lMeasure and source independently
lAutomated cold junction compensation
l6 digits, 320×240 bitmap graphics LCD screen
lEasy to operate
lWith RS232 interface
lDC/AC interchangeable, built-in rechargeable Ni-MH battery
lDimension :300×270×160 mm Net weight :3.5kg
Specifications
Temperature Measure (thermocouple sensors )
TC type | Measure Range (°C) | Stability(°C/year) | Resolution(°C) |
K | -200 ~ -100 | 1.0 | 0.01
|
-100 ~ 400 | 0.5 | ||
400 ~ 1370 | 0.4 | ||
E | -250 ~ -200 | 1.4 | |
-200 ~ -100 | 0.4 | ||
-100 ~ 600 | 0.25 | ||
600 ~ 1000 | 0.2 | ||
J | -210 ~ 100 | 0.7 | |
-100 ~ 800 | 0.35 | ||
800 ~ 1200 | 0.25 | ||
T | -250 ~ -200 | 2.0 | |
-200 ~ 0 | 0.9 | ||
0 ~ 400 | 0.4 | ||
N | -200 ~ -100 | 1.4 | |
-100 ~ 900 | 0.7 | ||
900 ~ 1300 | 0.4 | ||
R | -20 ~ 0 | 3.5 | |
0 ~ 100 | 2.8 | ||
100 ~ 1760 | 1.8 | ||
S | -20 ~ 0 | 2.8 | |
0~ 200 | 2.5 | ||
200 ~ 1400 | 1.8 | ||
1400 ~ 1760 | 1.4 | ||
B | 600 ~ 800 | 2.3 | |
800 ~ 1000 | 2.0 | ||
1000 ~ 1820 | 1.3 |
Temperature Measure (thermal resistors sensors )
RTD type | Measure Range(°C) | Stability(°C/year) | Resolution (°C) |
PT100-3 | -200~+850 | 0.6 | 0.01
|
PT100-4 | 0.5 | ||
PT10-3 | 2.0 | ||
PT10-4 | 1.5 | ||
Cu50-3 | -50~+150
| 0.7 | |
Cu50-4 | 0.5 | ||
Cu100-3 | 0.7 | ||
Cu100-4 | 0.5 |
DC Electricity Signal Measure
Signals | Measure Range | Accuracy | Resolution |
mV | (0~±200)mV | 0.01%R.D+0.005%F.S | 1uV |
V | (0~±30)V | 0.1mV | |
mA | (0~±25)mA | 0.1uA | |
R-3W | (0~400)Ω | 0.02%R.D+0.005%F.S | 1mΩ |
R-4W | (0~400)Ω | 0.015%R.D+0.005%F.S | 1mΩ |
Hz | (0~50)KHz | 0.01%R.D+0.005%F.S | 0.1Hz |
Remark: 1. The above inaccuracy does not include cold joint compensated temperature inaccuracy
Cold joint compensated temperature inaccuracy is ±0.5°C 2 . The above datas are valid in 22°C±5°C
3.The source information is abbreviated here, can be offered as per your need.
Multifunction Process Calibrator