- Contact Person : Ms. Shaw Ana
- Company Name : Beijing Spake Technology Co., Ltd.
- Tel : 86-10-6271 3169
- Fax : 010-62713168
- Address : Beijing,Beijing,F6,Jinyanlong Mansion,Xisanqi,Haidian,Beijing
- Country/Region : China
- Zip : 100096
SPMK3005 Multifunction Process Calibrator
Applications:
SPMK3005 Multifunction Process Calibrator (also named as: Thermometer Calibrator) is customized for calibration of thermodynamic instruments, pressure gauges etc.
Features:
lSource current, voltage, resistance, frequency, pressure.
lSimulate electrical signals output of thermocouple and thermal resistors.
lMeasure and source independently
lAutomated cold junction compensation
l6 digits, 320×240 bitmap graphics LCD screen
lEasy to operate
lWith RS232 interface
lDC/AC interchangeable, built-in rechargeable Ni-MH battery
lDimension :300×270×160 mm Net weight :3.5kg
Specifications
Temperature Measure (thermal resistors sensors )
RTD type | Measure Range(°C) | Stability(°C/year) | Resolution (°C) |
PT100-3 | -200~+850 | 0.6 | 0.01
|
PT100-4 | 0.5 | ||
PT10-3 | 2.0 | ||
PT10-4 | 1.5 | ||
Cu50-3 | -50~+150
| 0.7 | |
Cu50-4 | 0.5 | ||
Cu100-3 | 0.7 | ||
Cu100-4 | 0.5 |
DC Electricity Signal Measure
Signals | Measure Range | Accuracy | Resolution |
mV | (0~±200)mV | 0.01%R.D+0.005%F.S | 1uV |
V | (0~±30)V | 0.1mV | |
mA | (0~±25)mA | 0.1uA | |
R-3W | (0~400)Ω | 0.02%R.D+0.005%F.S | 1mΩ |
R-4W | (0~400)Ω | 0.015%R.D+0.005%F.S | 1mΩ |
Hz | (0~50)KHz | 0.01%R.D+0.005%F.S | 0.1Hz |
Temperature Measure (thermocouple sensors )
TC type | Measure Range (°C) | Stability(°C/year) | Resolution(°C) |
K | -200 ~ -100 | 1.0 | 0.01
|
-100 ~ 400 | 0.5 | ||
400 ~ 1370 | 0.4 | ||
E | -250 ~ -200 | 1.4 | |
-200 ~ -100 | 0.4 | ||
-100 ~ 600 | 0.25 | ||
600 ~ 1000 | 0.2 | ||
J | -210 ~ 100 | 0.7 | |
-100 ~ 800 | 0.35 | ||
800 ~ 1200 | 0.25 | ||
T | -250 ~ -200 | 2.0 | |
-200 ~ 0 | 0.9 | ||
0 ~ 400 | 0.4 | ||
N | -200 ~ -100 | 1.4 | |
-100 ~ 900 | 0.7 | ||
900 ~ 1300 | 0.4 | ||
R | -20 ~ 0 | 3.5 | |
0 ~ 100 | 2.8 | ||
100 ~ 1760 | 1.8 | ||
S | -20 ~ 0 | 2.8 | |
0~ 200 | 2.5 | ||
200 ~ 1400 | 1.8 | ||
1400 ~ 1760 | 1.4 | ||
B | 600 ~ 800 | 2.3 | |
800 ~ 1000 | 2.0 | ||
1000 ~ 1820 | 1.3 |
Remark:1. The source information is abbreviated here, can be offered as per your need.
2. The above inaccuracy does not include cold joint compensated temperature inaccuracy
Cold joint compensated temperature inaccuracy is ±0.5°C
3. The above datas are valid in 22°C±5°C
SPMK3005 Multifunction Process Calibrator